• Biological safety cabin type: CLASS II according to EN 12469 regulation.
• Work area and work surface built in stainless steel AISI 304.
• Sliding front glass with alarm position. Automatic or Manual.
• Glass side walls.
• The recycled air is filtered by an absolute ULPA U15 filter (efficiency 99.9995% MPPS Standard EN1822) and sent to the work area at a speed of 0.45 m/s
• Partial air recirculation according to regulations: 30% ejection and 70% recirculation.
• The air expelled to the outside is filtered by an absolute filter ULPA U15 (efficiency 99.9995% MPPS Standard EN1822) independent dedicated to this purpose.
• Work area maintains laminar flow conditions with an air purity type ISO 4 (According to ISO 14644) or Class 10 (according to Fed. Std. 209).
• Protection barrier in work area with speed greater than 0.53 m / s.
• Microprocessor-controlled cab with upgradeable FLASH memory and external color touch display for easy visualization of cabin control parameters (speeds, flow rates, plenum pressure, ejection barriers, humidity, temperature, etc ...).
• The cab is easy to operate for white/ultraviolet lighting control, internal electrical outlet.
• Alarms to detect problems in the drive system and the protection barrier by negative pressure and door opening, by voice and light warning on the screen.
• Fluorescent and UV lighting excluded.
• Working voltage 110-220 V. 50/60Hz. Maximum current consumption 4 A. (approx. 800 w.).
• Standard cabin Includes U.V. light; 2 Electrical outlets of 110-220 volts.
Optional Accessories:
• Gas intake with electronically controlled valve.
• Vacuum socket.
• Hydrogen Peroxide decontamination connector.